The fluorescence detector RF-20Axs provides world-class sensitivity, excellent maintainability and diverse validation / support functions. It supports a wide range of applications in the wavelength range of 200 to 750 nm from conventional analysis to high-performance analysis.With a signal-to-noise ratio of 2000 for the water-Raman band, the fluorescence detector is very well suited for trace analysis.Measuring cell tempering ensures the lowest possible noise and drift values
The integrated low-pressure mercury lamp allows easy wavelength accuracy verification. The xenon lamp and flow cell are directly accessible on the device, thus allowing a quick and easy handling and maintenance of the device by the user, thereby minimizing downtime. The lamp life is 2000 hours. When replacing the xenon lamp, no adjustment is required.
Technical Data
Detector type |
Fluorescence detector |
Detection channels |
2 |
Number of signals |
2 |
Light source |
xenon lamp, low-pressure; mercury lamp for wavelength accuracy check |
Wavelength range |
200 – 750 nm |
Spectral bandwidth |
20 nm |
Wavelength accuracy |
± 2 nm |
Wavelength precision |
± 0.2 nm Indicates the precision performance when the power is turned ON in single wavelength mode and the wavelength is changed. |
Sensitivity |
can be set at three levels: HIGH (x 1), MED (x 32), LOW (x 1024) |
Wetted materials |
SUS316L, PTFE (fluorocarbon polymers), quartz |
Flow cell volume |
12 µl |
Time constants |
11 levels can be selected, equivalent to “no filter”, 0.05, 0.1, 0.5, 1.0, 1.5, 2.0, 3.0, 6.0, 8.0 and 10.0 seconds |
Autozero |
auto zero function, basline shift function |
Temperature range |
room temperature (ambient temperature) -10 °C up to 40 °C (at flow rates less than 2 ml/min, column oven temperatures less than 85 °C) |
Gain |
can be set at three levels: x 1, x 4, x 16 |
Ambient conditions |
operating temperature: 4 to 35 °C, relative humidity: 20 to 85 % (no condensation) |
Power supply |
AC220–240 V, 400 VA, 50/60 Hz |
Dimensions |
260 x 210 x 420 mm (W x H x D) |
Weight |
18 kg |
-
Fluoreszenz-Detektor RF-20A/20Axs Benutzerhandbuch DE
-
Fluorescence Detector RF-20A/20Axs User Manual EN
-
List of supported devices in chromatography data systems EN (V1663)
Device support in KNAUER CDS / KNAUER device support in 3rd party software