Load your samples and press the START button. The tray rotates and the arm with the needle is aligned on top of the first sample vial to be processed.
The needle is inserted in the vial at the programmable depth where the sample is aspirated by the built-in pump in order to fill the loop. The injection valve is then moved to the inject position and the sample is delivered to the HPLC system. Then, by using the on-board solvents, the needle and the whole flow-path are cleaned before processing the next sample.
HT1500L/ – uses standard consumables
– offers the capability to automate up to 45 samples. By supporting multiple racks – each of them removable – allows for continuous feeding and convenient operation.
Keypad operations enable simple automation of any existing HPLC and makes the autosampler easy to control by any chromatography software.
HT1500L comes with HTA Autosampler Manager (Standard Version), our complimentary software control for autosampler programming. After setting your method once, just load samples and press START
Is your HPLC routine calling for automation? Look no further than the HT1500L; an autosampler designed for HPLC, fully equipped with powerful features.
Unlike LC-MS, nanoLC and UHPLC systems the HPLC system requires few, essential features: excellent analytical performance, good sample capacity, and of course, uncompromised reliability, robustness and durability.
We have designed the HT1500L with all essential features a cutting-edge HPLC system calls for, while eliminating anything unnecessary to keep your costs as a minimum. No frills. No compromise.
While we have struggled to achieve a low price, we haven’t made any compromise on quality.
HT1500L has a strong commitment to quality that focuses on providing a real unattended experience: HT1500L will be your workhorse requiring very few or no attention from lab technicians. The materials have been picked up to provide longevity; the up-to-date technologies offer reliability that deserve a modern instrument while the design attests our care in details.
Technical Specifications
Sample capacity: |
45 vials, 4ml (5 removable racks) |
Solvent capacity: |
5 vials, 40ml (1 removable rack) |
Maintenance: |
Preventive counters available |
Electrical control: |
USB and TTL |
RSD: |
≤ 0.25% |
Linearity (r): |
≥ 0.999 |
Carry over1: |
≤ 0.01% |
Mode: |
every injection/sample/step pre & post injection |
Sampling depth: |
programmable |
Filling speed: |
programmable |
Viscosity delay: |
0-25s |
Injection mode: |
Full Loop |
Supported loops |
20; 50; 100ml |
Cycle time: |
≤ 60sec |
Analysis time: |
0-999min |
Dimensions (WxHxD)2: |
355x380x520mm |
Weight: |
8kg |
Power supply: |