Preparative HPLC pump
AZURA preparative HPLC pump P 2.1L covers wide flow rate range and pressure capabilities. It has been designed for purification of mg and gram samples. The Mobile Control allows intuitive control in stand-alone operation. The integrated automatic recognition of the pump head with RFID technology allows fast adaptions of the pump for various applications.
Technical Data
Control |
LAN; RS-232; analog and event controlled |
Inputs |
flow rate |
Outputs |
pressure; 8 event outputs (TTL, OC, relays) |
Analog inputs |
0-10 V |
Analog outputs |
1, 2 or 5 V |
Power supply |
100 – 240 V, 50 – 60 Hz |
Dimensions |
361 x 208,2 x 523 mm (B x W D) |
Weight |
19 kg |
Leak sensor |
yes |
Ambient conditions |
temperature range: 10–40 °C; 50–104 °F
air humidity: below 90 % humidity (non condensing) |
Max. flow rate |
500 ml/min |
Flow rate increment |
0.01 ml/min |
Maximum delivery pressure [psi] |
1450 psi |
Maximum delivery pressure [bar] |
100 bar |
Maximum delivery pressure [MPa] |
10 MPa |
Pump head materials |
Titanium |
Maximum viscosity |
1000 cP |
Liquid temperature range |
4−60 °C (39.2−140 °F) |
Gradient |
isocratic |
Leak management |
yes |
Wetted materials |
Titanium, aluminum oxide (Al2O3), FKM, GFP (graphite fiber reinforced PTFE), PEEK, ruby, sapphire |
Note |
Pressure sensor membrane made of stainless steel; limited wetting in fluidic path |
Best working conditions |
1-400 ml/min |
Continous working conditions |
2-200 ml/min |
Flow rate accuracy |
± 2 % |
Flow rate accuracy conditions |
measured at 5 – 50 % of flow range using ethanol/water 10:90 |
Flow rate precision |
< 0.1 % RSD |
Flow rate precision conditions |
measured at 5 – 50 % of flow range using ethanol/water 10:90 |
HPG: gradient accuracy |
± 2 % (5 – 95 %, measured at 5 – 50 % of flow range, water/caffeine tracer) |
HPG: gradient precision |
1 % RSD, based on retention time at constant room temperature |
LPG: gradient accuracy |
± 3 % (5 – 95 %, measured at 5 – 50 % of the flow range, water/caffeine tracer) |
LPG: gradient precision |
2 % RSD, based on retention time at constant room |
Active piston seal wash |
+ |
Piston seal washing |
+ |
Pump head inlet (standard) |
M8 x 1 (flat bottom) |
Pump head outlet (standard) |
M8 x 1 (coned) |
-
AZURA Pumpe P 2.1L Benutzerhandbuch DE (V6840)
-
List of supported devices in chromatography data systems