AZURA P 6.1L

Product No:

APH35ED

Binary High pressure gradient 10 ml / Min S.S. With degasser for normal phase applications


AZURA P 6.1L High Pressure Pump with 10 ml pump head for Normal Phase Applications, stainless steel

Technical Data

General Information

Max. flow rate 10 ml/min
Flow rate range 0.001 – 10 ml/min
Flow rate increment 0.001 ml/min
Maximum delivery pressure [psi] 10150 psi
Maximum delivery pressure [bar] 700 bar
Maximum delivery pressure [MPa] 70 MPa
Pressure range
Pump head materials Stainless Steel
Gradient High Pressure Gradient (binary)
Leak management Yes
Wetted materials GFP (graphite fiber reinforced PTFE), sapphire, ruby, aluminum oxide (Al2O3)

Detailed information

Best working conditions 0.1 – 8.0 ml/min
Continous working conditions 0.1 – 4.0 ml/min
Flow rate accuracy < 1 %
Flow rate accuracy conditions measured at 5-80 % of flow range, using ethanol
Flow rate precision 0.1 % RSD
Flow rate precision conditions based on the retention time at constant room temperature
Gradient range 0 – 100%
Gradient range (recommended) 5 – 95 %
Gradient accuracy ± 0.3 % at 1 ml/min, 150 bar (ethanol/caffeine tracer) ± 1 % (5 – 95 %, measured at 0.1 – 10 ml/min, water/caffeine tracer)
Gradient precision < 0.1 % RSD at 1 ml/min, 0.3% RSD overall, based on retention time at constant room temperature
Gradient minimum increment 0.1 %
Mixing volume 100 µl
Delay volume 110 µl with 100 µl Mixer
Pump type Binary analytical HPLC pump with degasser
Degasser Yes
Solvent selection valve 2 x 2 channels
Pulsation < 2 % Amplitude (typically: < 1,3 %) or < 3 bar (0,3 MPa), whatever is greater, at 1 ml/min ethanol, at all pressures > 10 bar (1 MPa, 147 psi)
Pulsation compensation Active pressure and pulsation compensation
Active piston seal wash Yes
Piston seal washing Yes
System protection Soft start, Pmin und Pmax are programmable
Pump head inlet (standard) UNF 1/4-28 Thread
Pump head outlet (standard) UNF 10-32 Thread

Communication

Interfaces LAN, Pin header connectors (Analog IN, Start In, Error IN)
Control LAN, Analog and event control, Mobile Control
Analog inputs Flow rate, 0 – 10 V via pin header connectors
Analog outputs 8 event outputs (TTL, OC, Relais) and 24 V

General

Power supply 100 – 240 V; 50 – 60 Hz; Maximum power consumption 100 Watt
Dimensions 361 mm x 208.2 mm x 523 mm (W × H × D)
Weight 14.1 kg
Leak sensor Yes
Ambient conditions 4-40 °C (39.2-104 °F) Air humidity below 90%, non-condensing

Downloads and more information

  • AZURA Pump P 6.1L User Manual EN (V6890)

    • Manual
    • English
    • PDF
    • 1.1MB
  • AZURA Pumpe P 6.1L Benutzerhandbuch DE (V6890)

    • Manual
    • German
    • PDF
    • 1.1 MB