Variable single wavelength with fiber optic connectors
The AZURA UV/VIS Detector UVD 2.1L is a competitively priced HPLC spectrophotometer for routine HPLC applications including fast LC methods. Besides offering excellent technical specifications, this robust detector features a highly flexible and compact design. The UVD 2.1L comes with an installed deuterium lamp which covers a wavelength range from 190 to 750 nm. This detector can be controlled with OpenLab® EZChrom Edition, ChromGate® and ClarityChrom® software, as well as from an optional touch display (stand-alone operation), via LAN or through analog input/output; allowing it to be integrated into almost any LC system. Due to a smart design the flow cell is easily accessible and can be changed very quickly. You can choose between a wide range of flow cells for analytical or preparative LC applications with flow rates from 10 μl/min up to 10 l/min. This fiber optics version offers the possibility to separate the flow cell spatially from the device and thus provides enhanced security for hazardous, explosive or toxic work processes.
Technical Data
Detector type |
Variable single wavelength UV detector |
Version |
fiber optics version |
Detection channels |
1 |
Light source |
Deuterium (D²) lamp with integrated GLP chip |
Wavelength range |
190 – 750 nm |
Spectral bandwidth |
11 nm at Hα line (FWHM) |
Wavelength accuracy |
± 2.5 nm |
Wavelength precision |
0.3 nm (ASTM E275-93) |
Noise |
± 2.0 x 10-5 AU at 254 nm (ASTM E1657-98) |
Drift |
4.0 x 10-4 AU/h at 254 nm (ASTM E1657-98) |
Linearity |
> 2.0 AU at 270 nm (ASTM E1657-98) |
Maximum data rate |
50 Hz (LAN)/20 Hz (Analog) |
Flow cell |
Not included (see Accessories / Spare parts) |
Time constants |
0.0 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s |
Integration time |
Automatic |
Leak sensor |
Yes |
Inputs |
Error (IN), Start (IN), Autozero,0 – 10 V Analog IN |
Outputs |
Events 1 – 3, + 5 V, 24 V Valve |
Analog outputs |
1 x 0 – 5 V scalable, 20 bit, offset adjustable |
Control |
Digital: LAN, remote connector/Analog: wavelength control/Manual: Mobile Control (optional) |
Programming |
Timed: wavelength, events, fraction valve, links, wake up (program, link); /9 programs, 50 program lines |
GLP |
Detailed report including lamp recognition, operating hours, /lamp operating hours, number of lamp ignitions |
Display |
Mobile Control (optional) |
Ambient conditions |
Temperature range: 4 – 40 °C, 39.2 – 104 °F, Humidity: below 90 % noncondensing |
Power supply |
100 – 240 V, 50 – 60 Hz, 75 W |
Dimensions |
361 x 158 x 523 mm (W x H x D) |
Weight |
6.0 kg |
Note |
Fiber optic cables and flow cells are not included and must be ordered/ separately (see Accessories / Spare parts) |
-
AZURA Detektor UVD 2.1L Benutzerhandbuch DE (V6831)
-
AZURA Detector UVD 2.1L User Manual EN (V6831)
-
List of supported devices in chromatography data systems EN (V1663)
Device support in KNAUER CDS / KNAUER device support in 3rd party software