with 1000 ml pump head, stainless steel
The Preparative Pump 80P delivers stable and precise solvent flow for semi-preparative as well as preparative HPLC applications. By adding one or more additional pumps, binary, ternary or even quaternary high-pressure gradient systems can be easily configured. The pump can also be equipped with a binary or ternary gradient valve block to configure cost-effective low-pressure gradient systems. Exchangeable pump heads for flow rates up to 1 000 ml/min and pressures up to 40 MPa make it possible to easily optimize the pump to changing performance requirements and allow simple maintenance. Available in stainless steel or biocompatible titanium versions, the pump heads enable a wide range of applications.
Technical Data
Control |
LAN; RS-232; analog and event controlled |
Inputs |
flow rate |
Outputs |
pressure; 8 event outputs (TTL, OC, relays) |
Analog inputs |
0-10 V |
Analog outputs |
1, 2 or 5 V |
Power supply |
100 – 240 V, 50 – 60 Hz |
Dimensions |
361 x 208,2 x 523 mm (B x W D) |
Weight |
19 kg |
Leak sensor |
no |
Ambient conditions |
temperature range: 10–40 °C; 50–104 °F
air humidity: below 90 % humidity (non condensing) |
Max. flow rate |
1000 ml/min |
Flow rate range |
1-1000ml |
Flow rate increment |
0.1 ml/min |
Maximum delivery pressure [psi] |
730 psi |
Maximum delivery pressure [bar] |
50 bar |
Maximum delivery pressure [MPa] |
5 MPa |
Pump head materials |
Stainless Steel |
Maximum viscosity |
1000 cP |
Liquid temperature range |
4−60 °C (39.2−140 °F) |
Gradient |
Isocratic |
Leak management |
– |
Wetted materials |
stainless steel, aluminum oxide (Al2O3), FKM, PEEK, ruby, sapphire, GFP (graphite fiber reinforced PTFE) |
Best working conditions |
2-800 ml/min |
Continous working conditions |
2-400 ml/min |
Flow rate accuracy |
± 2 % |
Flow rate accuracy conditions |
measured at 5 – 50 % of flow range using ethanol/water 10:90 |
Flow rate precision |
< 0.1 % RSD |
Flow rate precision conditions |
measured at 5 – 50 % of flow range using ethanol/water 10:90 |
HPG: gradient accuracy |
± 2 % (5 – 95 %, measured at 5 – 50 % of flow range, water/caffeine tracer) |
HPG: gradient precision |
1 % RSD, based on retention time at constant room temperature |
LPG: gradient accuracy |
± 3 % (5 – 95 %, measured at 5 – 50 % of the flow range, water/caffeine tracer) |
LPG: gradient precision |
2 % RSD, based on retention time at constant room |
Active piston seal wash |
+ |
Piston seal washing |
+ |
Pump head inlet (standard) |
M8 x 1 (flat bottom) |
Pump head outlet (standard) |
M8 x 1 (coned) |
-
AZURA Pumpe P 2.1L Benutzerhandbuch DE (V6840)
-
List of supported devices in chromatography data systems EN (V1663)
Device support in KNAUER CDS / KNAUER device support in 3rd party software